A CBCT (Cone Beam Computed Tomography) transformed the three-dimensional assessment of dentomaxillofacial structures. Unlike two-dimensional radiographs, it allows you to visualize structures in different planes and analyze anatomical relationships that may be overlapping or poorly defined in conventional images.
But there is an important difference between having a three-dimensional image and having an adequate three-dimensional image to answer clinical question.
The quality of a CBCT does not simply depend on choosing “the highest resolution possible.” It results from the combination of indication, field of view, voxel size, exposure parameters, patient positioning, movement, reconstruction and interpretation.
The indication of CBCT must start from a question.
It is not:
“Is it worth doing a CT scan?”
It is:
“What diagnostic information do I need to obtain that clinical evaluation and conventional exams cannot adequately provide?”
This change in reasoning is fundamental.
CBCT can be particularly useful when three-dimensional information influences diagnosis or planning, as in certain cases of implant dentistry, endodontics, impacted teeth, evaluation of anatomical structures, and orthodontic-surgical planning.
This does not mean that every patient should undergo CBCT.
The referral needs to be individually justified, considering diagnostic benefit and radiation exposure. This is also the position of entities such as the American Academy of Oral and Maxillofacial Radiology (AAOMR).
FOV (Field of View) is the anatomical region included in the acquisition. And here is an extremely important rule:
The FOV should be sufficient to answer the clinical question-not larger than necessary.
If the goal is to assess a localized region, it makes no sense to acquire a much larger area just because the equipment allows it.
Reducing the FOV may contribute to reducing radiation exposure. Studies demonstrate a relationship between field size and dose, with relevant reductions when the field is limited to the region of interest.
In addition, a smaller FOV may be advantageous for certain applications because it focuses acquisition on the region that actually needs to be analyzed.
More area does not necessarily mean more useful information.
No.
O voxel is the smallest volumetric unit that makes up the three-dimensional image. In a simplified way: smaller voxel → potentially higher spatial resolution.
But this does not mean that we should always choose the smallest voxel available. The voxel size must be compatible with the diagnostic need.
A systematic review on different voxel sizes showed that the relationship between voxel and diagnostic performance cannot be summarized as “smaller is better”. The benefit of higher resolution depends on the diagnostic task and other protocol factors.
This is particularly important because increasing resolution can have implications on noise, dose, and acquisition time, depending on the protocol.
Therefore:
Choose the resolution needed for the clinical question — not the maximum resolution just because it exists.
The exposure parameters directly affect the quantity and quality of the photons used to form the image.
In simplified form:
kV (kilovoltage) is related to the energy of the X-ray beam.
mA (milliamp) is related to the amount of radiation produced per unit time.
The product between current and exposure time also influences the total amount of radiation used.
But there is no universal combination of values that works for all patients and all indications.
The protocol needs to consider anatomical region, patient size, indication and required image quality.
A review on dose optimization highlights that the selection of parameters should seek the lowest exposure compatible with a diagnostically acceptable image.
This is one of the pitfalls of CBCT.
A visually impressive three-dimensional reconstruction does not necessarily mean that all details are faithfully represented. There are artifacts that can interfere with the interpretation.
Among them are changes related to:
patient movement;
metallic objects;
beam hardening;
spread;
detector limitations;
reconstruction process.
These artifacts can produce structures or changes that do not exactly match the real anatomy.
Therefore, the professional needs to recognize when an image may be technically limited before turning it into a clinical conclusion.
During acquisition, the equipment needs to record multiple projections to reconstruct the volume.
If the patient moves, these projections may no longer represent the exact same anatomical position.
The result may be: blurring + duplication + loss of definition + motion artifacts.
This seems basic, but it is an important variable because there is no further processing capable of perfectly retrieving information that has not been acquired correctly.
Positioning is also part of diagnostic quality.
When we talk about quality, we are talking about a set of characteristics.
Among them:
spatial resolution
ability to distinguish small and close structures;
contrast
ability to differentiate structures with different attenuation characteristics;
noise
random variations that may impair visualization;
artifacts
information present in the image that does not adequately correspond to the actual object.
Spatial resolution is particularly relevant in dentistry because many applications rely on the visualization of small structures. However, it is influenced not only by the nominal voxel size, but also by the detector, reconstruction, motion, and other parameters of the system.
In implant dentistry, three-dimensional information makes it possible to evaluate aspects such as:
height and width of the ridge;
relationship with anatomical structures;
bone morphology;
position of the mandibular canal;
relationship with maxillary sinus;
characteristics of the receptor site.
A systematic review on implant planning demonstrated that the linear measurements obtained by CBCT show good correspondence with clinical and experimental references, although there are limitations and variations according to the protocol and the region evaluated.
More recently, a systematic review and meta-analysis showed that low-dose protocols can provide adequate precision for certain measurements used in implant planning, as long as the necessary diagnostic quality is preserved.
This brings up an important conclusion:
The goal is not to obtain the most detailed image possible. It's about getting the right image to make a clinical decision.
CBCT can add three-dimensional information in situations where conventional exams do not adequately answer the diagnostic question.
This may be particularly relevant in complex situations, such as:
complex root anatomy;
suspicion of additional channels;
resorptions;
root fractures;
periapical lesions;
planning of surgical procedures.
But the indication should remain guided by clinical need.
The 2025 joint update of the American Association of Endodontists (AAE) and AAOMR reinforces precisely the need for selective use of CBCT in endodontics, based on the clinical situation and the information to be obtained.
In orthodontics, CBCT can provide relevant three-dimensional information in selected cases, especially when there are questions that cannot be adequately answered by conventional images.
However, it should not be requested simply as an automatic part of the documentation.
The nomination must consider:
which question needs to be answered?
does three-dimensional information change the diagnosis?
does it modify the planning?
benefit justifies exposure?
This principle of individualized indication is central to AAOMR's recommendations for the use of CBCT in orthodontics.
There is an idea that needs to be very clear:
The best CBCT is not the one that uses the highest dose. It is the one that delivers sufficient diagnostic information with the least exposure reasonably possible.
The optimization principle should accompany the indication.
This involves:
justify the exam → limit the FOV → select appropriate protocol → avoid unnecessary repetitions → preserve diagnostic quality.
The literature shows great dose variation between equipment and protocols, which means that it is not appropriate to talk about a universal “standard dose” for dental CBCT.
Producing the volume is not the end of the process.
Interpretation should consider the set of images and the different reconstruction planes.
A three-dimensional evaluation may involve:
axial + coronal + sagittal + panoramic reconstructions + specific reconstructions, as indicated.
And there is an important caution:
Do not interpret only the image that confirms the initial hypothesis.
The assessment should be systematic, especially when the volume covers structures beyond the region that motivated the examination.
In the end, we can summarize it in six questions:
1. Why am I requesting the exam?
2. Which region do I need to visualize?
3. Which FOV is sufficient?
4. What resolution is needed?
5. Which protocol offers adequate diagnostic quality with lower exposure?
6. Is the image obtained technically adequate to answer the clinical question?
If these six questions are well answered, you are not simply “doing a CT scan”.
It is building a diagnostic-driven acquisition.
CBCT should not be understood as a “more complete” version of conventional radiography.
It is a three-dimensional modality with advantages, limitations, biological costs, and interpretation needs of its own.
Its greatest value appears when three-dimensional information actually adds something to the clinical decision.
More images do not necessarily mean more diagnosis. A well-indicated, well-acquired, and correctly interpreted image is much more valuable.
SCHULZE, R. K. W.; DRAGE, n. A. Cone-beam computed tomography and its applications in dental and maxillofacial radiology. Clinical Radiology, v. 75, n. 9, p. 647-657, 2020. DOI: 10.1016/j.crad.2020.04.006.
VAN DOREN, C. et al. Technical aspects of dental CBCT: state of the art. Dentomaxillofacial Radiology, 2014. DOI: 10.1259/dmfr.20140224.
SPIN-NETO, Rubens; GOTFREDSEN, Erik; WENZEL, Ann. Impact of voxel size variation on CBCT-based diagnostic outcome in dentistry: a systematic review. International Journal of Computer Assisted Radiology and Surgery, v. 8, n. 4, p. 813-820, 2013. DOI: 10.1007/s11548-012-0735-9.
FOKAS, George et al. Accuracy of linear measurements on CBCT images related to presurgical implant treatment planning: a systematic review. Clinical Oral Implants Research, v. 29, suppl. 16, p. 393-415, 2018. DOI: 10.1111/clr.13142.
CARNEIRO, Ana Luiza E. et al. Accuracy of linear measurements for implant planning based on low-dose cone beam CT protocols: a systematic review and meta-analysis. Dentomaxillofacial Radiology, v. 53, n. 4, p. 207-221, 2024. DOI: 10.1093/dmfr/twae007.
SCHULZE, R. et al. Artefacts in CBCT: a review. Dentomaxillofacial Radiology, v. 40, n. 5, p. 265-273, 2011. DOI: 10.1259/dmfr/30642039.